AquaC UNO H [120V, 60Hz, 1700VA, US-Version- a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-0970-1 Vivonic Part Number: G02000966-US

Company
Fresenius Medical Care Holdings, Inc.
Recall Initiated
June 21, 2022
Posted
September 19, 2022
Recall Number
Z-1770-2022
Quantity
6 units
Firm Location
Waltham, MA

Reason for Recall

Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

Distribution

US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.

Lot / Code Info

UDI: (01)4251285200234(11)200422(21)0CHS3987 (01)4251285200234(11)200526(21)0CHS4108 (01)4251285200234(11)210623(21)1CHS5372 (01)4251285200234(11)210707(21)1CHS5405 (01)4251285200234(11)190305(21)9CHS2881 (01)4251285200234(11)190529(21)9CHS3090 Serial Numbers: 0CHS3987 0CHS4108 1CHS5372 1CHS5405 9CHS2881 9CHS3090

Root Cause

Under Investigation by firm

Action Taken

Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) issued Urgent Medical Deice Correction letter (FA-2022-20-C) on 06/21/22 via First Class Mail. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) Water Systems Service Specialist will be contacting you to schedule a correction of your portable Water Treatment system. Only a Certified AquaC UNO H System Technician or FMCRTG Water Systems Service Specialist may perform this service. If choosing to correct via your own Certified AquaC UNO H System Technician, you will need to contact FMCRTG prior to performing the correction. FMCRTG will provide the necessary instruction via a Change Notification. For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.