Roadrunner PC Wire Guide (Nimble Floppy), Guidewire RFPC-35-145 G07914

Company
Cook Incorporated
Recall Initiated
August 16, 2022
Posted
September 28, 2022
Recall Number
Z-1839-2022
Quantity
202 US; 11 OUS
Firm Location
Bloomington, IN

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

Lot / Code Info

RFPC-35-145 G07914 UDI-DI: 00827002079144 Lots/Expiration Date: 14828624 06-07-2025 14828638 06-07-2025 14837046 12-07-2025 NS14828650 06-07-2025

Root Cause

Under Investigation by firm

Action Taken

On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235