OPTI SARS-CoV-2 RT-PCR Test, REF Numbers: a) 99-57003 b) 99-57004

Company
Opti Medical Systems Inc
Recall Initiated
June 25, 2021
Recall Number
Z-2310-2021
Quantity
3552 kits
Firm Location
Roswell, GA

Reason for Recall

The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA Master Mix.

Distribution

Maine, New York, Alabama, Pennsylvania, Florida, Illinois, New Jersey, Connecticut, Georgia, Puerto Rico, Costa Rica, Hong Kong, Philippines, Barbados

Lot / Code Info

a) 99-57003, Lot Numbers: 20808, 20812, 20813 b) 99-57004, Lot Numbers: 20881, 20901, 211013

Root Cause

Nonconforming Material/Component

Action Taken

/The firm disseminated a technical bulletin beginning on 06/25/2021 by email identifying the /problem and providing additional instructions to be implemented when using the product.