Medline Tissue Recovery Instrument Set, Reorder #DYNJ64347A, packaged 3/case, Sterile, Rx. Convenience kit used to recover tissue, which can be used for medical research, education, or transplantation.

Company
Medline Industries Inc
Recall Initiated
April 30, 2021
Recall Number
Z-1841-2021
Quantity
36/3-kit cases (108 kits)
Firm Location
Northfield, IL

Reason for Recall

The devices may not meet the minimum required sterility assurance level.

Distribution

US Distribution to state of: WI.

Lot / Code Info

GTIN (01)10193489291186, lot numbers 21CDA959, 21BDB665, 21ADA390, and 20KDC285

Root Cause

Software change control

Action Taken

The recalling firm, Medline Industries, Inc., issued an "IMMEDIATE ACTION REQUIRED RECALL" letter dated and emailed on 4/30/2021. The letter described the product, the reason for recall and actions to be taken. The consignees were instructed to immediately check their inventory for the affected product, complete and submit response form online to: Website link: https://recalls.medline.com; Recall Reference #: R-21-073; Recall Code: R5A2C566 , and quarantine it for subsequent return. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, Inc. If you have any questions, please contact the Recall Department at 866-359-1704.