The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

Company
Canon Medical System, Usa, Inc.
Recall Initiated
November 15, 2023
Posted
February 16, 2024
Recall Number
Z-1154-2024
Quantity
1
Firm Location
Tustin, CA

Reason for Recall

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Distribution

US: CA, LA

Lot / Code Info

UDI-DI: 04987670100192, Serial Number: B1E22X2005

Root Cause

Component change control

Action Taken

On 11/15/23 correction notices were distributed to customers who were asked to do the following: 1) Should this issue occur please attempt to restart the system and contact your service representative. 2) A recalling firm representative will contact you to schedule a time to replace the flat panel detector on the console. The parts required to address the defect are expected to be available no earlier that December 2023. 3) Complete and return the customer reply form via email to RAFFAIRS@US.MEDICAL.CANON For questions, please contact our InTouch team at intouchdispatch@us.medical.canon or 800-421-1968, or 800.521.1968.

More recalls by Canon Medical System, Usa, Inc.

View all recalls by this company →