RecallDepth

Brasseler 8S RA SHP RD (Super-Sharp Round) H1S.21.023, REF 001156T0), carbide dental bur, 10 burs/package, Rx.

This recall is currently active, issued February 7, 2024. It was issued by Brasseler Usa I Lp.

Company
Brasseler Usa I Lp
Recall Initiated
January 11, 2024
Posted
February 7, 2024
Recall Number
Z-1008-2024
Quantity
440 burs
Firm Location
Savannah, GA
Official Source
View on FDA website ↗

Reason for Recall

The device has a grip detail (right angle latch) out of specification causing the bur to not adequately latch in the handpiece.

Distribution

Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, IL, NY, OH, OR, PA, TN, TX, and WA. The country of Canada.

Lot / Code Info

Lot numbers L01G6 and L16P9, UDI-DI 00887919188483, RX.

Root Cause

Under Investigation by firm

Action Taken

The recalling firm issued letters dated 1/8/2024 on 1/11/2024 via certified mail. The letter explains the reason for recall and required customer actions. The actions included that upon receipt of the letter, confirm the quantity of unused recalled product in their inventory and set it aside for return. Return information will be emailed to the consignee with a return label. The Customer Service emailed the letters to consignees on 1/12/2024 and sent product return instructions in the email. Phone calls will be made on 1/16/2024 if no response if received from the consignee via email. The recalling firm will be recording the amount of product to be returned on the phone log which will be completed for every consignee contacted. During the call, the consignees will also be given the product return instructions.

More recalls by Brasseler Usa I Lp

View all recalls by this company →