RecallDepth

CardioQuip Modular Cooler-Heater, Model: MCH-1000(m), Power: 115V-60Hz, 16A

Company
Cardioquip, Llc
Recall Initiated
October 24, 2023
Posted
December 15, 2023
Recall Number
Z-0566-2024
Quantity
2 units
Firm Location
College Station, TX

Reason for Recall

There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection.

Distribution

US: California, Texas, Pennsylvania, Washington, Ohio, Nebraska OUS: None

Lot / Code Info

UDI: 00860000846110 /Lot #:11160294 and 11161110

Root Cause

Process control

Action Taken

On October 24, 2023, CardioQuip issued a "Urgent: Medical Device Correction/Removal" notification via USPS to all affected consignees. CardioQuip ask consignees to take the following actions: 1. CardioQuip recommends that you remove your potentially contaminated MCH device from service immediately. 2. CardioQuip Customer Service will contact you by November 3, 2023 to provide your facility with a loaner device and to schedule an IWPR procedure for your MCH device. 3. Please complete and return the Receipt of Communication/Response Form attached to this letter to CardioQuip as soon as possible. 4. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program:

More recalls by Cardioquip, Llc

View all recalls by this company →