RecallDepth

BULK LARGEBORE MALE LL FITTING, Article No. M4341750SN. IV administration component.

This recall is currently active, issued November 3, 2023. It was issued by B Braun Medical Inc.

Company
B Braun Medical Inc
Recall Initiated
September 26, 2023
Posted
November 3, 2023
Recall Number
Z-0136-2024
Quantity
175,000 units
Firm Location
Allentown, PA
Official Source
View on FDA website ↗

Reason for Recall

Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous medication. There is also the potential for life-threatening situations to occur requiring additional medical intervention.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada and Singapore.

Lot / Code Info

UDI-DI: N/A Lot 0061744741, 0061749486

Root Cause

Process control

Action Taken

On September 26, 2023, the firm notified customers via Urgent Product Recall Notification letters. Customers were informed of the potential for leakage at the male Luer lock site. Customers were instructed to report total amount of impacted product on hand to the firm, after which B Braun Customer Support will contact the customer with instructions on how to return the product for credit or replacement depending on the customer's need. Questions about this recall may be reported to BBMI s Postmarket Surveillance Department by calling 1-833-425-1464.

More recalls by B Braun Medical Inc

View all recalls by this company →