VariSoft infusion set, Model Number 1002827

Company
Unomedical A/S Aaholmvej
Recall Initiated
October 11, 2023
Posted
November 21, 2023
Recall Number
Z-0232-2024
Quantity
8028 units

Reason for Recall

The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

Distribution

US Nationwide distribution in the state of CA.

Lot / Code Info

GTIN 05705244018709, Serial Numbers: 5388357, 5388362, 5388372, 5388366; GTIN 05705244018693, Serial Numbers: 5388357, 5388362, 5388372, 5388366

Root Cause

Process change control

Action Taken

Unomedical a/s issued an Urgent Field Safety Notice to its sole consignee on 10/11/2023 via email. The notice explained the issue, potential risk to patient, and requested the distributor to quarantine all affected product pending return and to notify their customers.