Intera 1.5T

This recall is currently active, issued November 21, 2023. It was issued by Philips North America Llc.

Company
Philips North America Llc
Recall Initiated
October 23, 2023
Posted
November 21, 2023
Recall Number
Z-0393-2024
Quantity
102 units
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

Potential component failure in the Gradient Coil could product smoke and/or fire.

Distribution

Domestic: Nationwide Distribution; Foreign: Albania, Andorra, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Libya, Luxembourg, Malaysia, Malta, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Russian Fed., Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, United States, UAE, Venezuela, Vietnam, & Yemen.

Lot / Code Info

Product No. 781195 & 781295; UDI-DI: N/A; Serial No. 18510, 18516, 18529, 18550, 18564, 18567, 18578, 18583, 18592, 18657, 18660, 18667, 18678, 18701, 18709, 18711, 18714, 18725, 18727, 18769, 18773, 18776, 18787, 18812, 18820, 18854, 18856, 18857, 18870, 18873, 18878, 18884, 18889, 18890, 18905, 18906, 18907, 18913, 18917, 18920, 18923, 18931, 18935, 18947, 18948, 18950, 18965, 18973, 18986, 18998, 20402, 20403, 20404, 20409, 20426, 20441, 30004, 30030, 30037, 30038, 30064, 30079, 30081, 30082, 30131, 30187, 30190, 30199, 30200, 30206, 30213, 30218, 30219, 30223, 30234, 30235, 30236, 30238, 30255, 30256, 30265, 30266, 30267, 30268, 30272, 30274, 30277, 30278, 30279, 30280, 30281, 30284, 30285, 30288, 30293, 30296, 30297, 30298, 30303, 30311, 30312, 30315.

Root Cause

Under Investigation by firm

Action Taken

The firm mailed URGENT Medical Device Correction Notices dated 10/23/2023 to consignees to notify them of this recall event. Consignees are instructed to identify if they possess any affected devices in their inventory and to follow the provided instructions for when scan aborts occur to prevent the failure mode associated with this recall event. Consignees are to stop the scan immediately and contact Philips Service if any of the failure situations outlined in the notice occur during a scan. Consignees are asked to complete and return the provided response form to Philips no later than 30 days after receipt of the recall notice. Questions can be directed to Customer Care Solutions at 1-800-722-9377.