RecallDepth

Alinity i Anti-TPO Reagent Kit, List Number 09P3521

Company
Abbott Ireland Limited Finisklin Business & Technology Park Sligo Ireland
Recall Initiated
April 24, 2023
Posted
June 8, 2023
Recall Number
Z-1696-2023
Quantity
110 kits

Reason for Recall

Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use (IFU).

Distribution

US Nationwide distribution in the states of CA, CO, FL, GA, IL, KY, LA, MD, MI, MN, NJ, NY, OH, RI, SC, SD, TN, TX, WA. No international distribution of affected lot.

Lot / Code Info

UDI Number: (01)00380740153892(17)240720(10)46009FN00; List Number: 09P3521; Lot Number: 46009FN00; Expiration Date: 07/20/2024

Root Cause

Nonconforming Material/Component

Action Taken

On April 24, 2023, Abbott issued a Product Correction letter, via FedEx priority overnight express, for this issue. Customers were instructed to provide a copy of the Product Correction letter to the laboratory manager, supervisor or health professional responsible for the impacted product and to complete and return the Customer Reply form.