Extended Tip Applicator 8CM, Box of 1.

Company
Integra Lifesciences Corp. (Neurosciences)
Recall Initiated
May 22, 2025
Posted
June 20, 2025
Recall Number
Z-1960-2025
Quantity
212 units
Firm Location
Princeton, NJ

Reason for Recall

Devices contain potentially out-of-specification levels of endotoxins and lack of sterility assurance.

Distribution

Domestic: Nationwide Distribution; Foreign: Canada, Japan, Australia, New Zealand, Taiwan (Province of China), Mexico, Colombia, Dominican Republic (the), Singapore, Mongolia, Hong Kong, Chile, Argentina, Puerto Rico, El Salvador, Guatemala, Korea (the Republic of), Peru, Costa Rica, Panama, From Belgium, distributed to: Austria, Belgium, Cyprus, Czechia, France, Georgia, Germany, Greece, Hungary, Iran (Islamic Republic of), Ireland, Italy, Jordan, Malta, Netherlands (the), Nigeria, Palestine, State of, Portugal, Romania, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Turkey.

Lot / Code Info

Model/Catalog Number: 201108; UDI 10381780000167; All Lots.

Root Cause

Process change control

Action Taken

An URGENT: VOLUNTARY MEDICAL DEVICE RECALL notification dated 5/22/25 was mailed to consignees for this product removal. Consignees are instructed to inspect their inventory for affected devices and quarantine any identified. The provided Acknowledgement form is to be completed and returned to Integra by email at FCA2@integralife.com or by FAX at 1-609-750-4220. The provided notification is to be forwarded to users of affected devices and a copy retained for consignee records. A Return Material Authorization number will be generated upon receipt of the Acknowledgment Form for credit. Sales Representatives and Distributors are instructed to inspect their inventory for affected devices and quarantine any identified. The provided Acknowledgement form is to be completed and returned to Integra by email at FCA2@integralife.com or by FAX at 1-609-750-4220. Distributors who further distributed affected devices are to forward the provided notification to their customers and collect Acknowledgement forms on their behalf to submit to Integra. Sales Representatives will be contacted by Post Market Quality to coordinate return of devices. Consignees with any questions are to contact Customer Service by phone at 1-800-654-2873 or by email at custsvcnj@integralife.com. Additional communications were made to customers on 7/10/25 and 7/18/25 informing them that expired devices are also subject to this recall and should be disposed of.

More recalls by Integra Lifesciences Corp. (Neurosciences)

View all recalls by this company →