RecallDepth

DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

Company
Deroyal Industries Inc
Recall Initiated
February 9, 2023
Posted
March 23, 2023
Recall Number
Z-1283-2023
Quantity
27,200 devices
Firm Location
Powell, TN

Reason for Recall

The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.

Distribution

US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.

Lot / Code Info

GTIN 50749756377046, Lot 58447585 exp. 12/22/2027

Root Cause

Process control

Action Taken

The firm issued an URGENT DEROYAL RECALL NOTICE its consignees by email and letter beginning on 02/09/2023. The notice explained the problem and risk, and requested the following actions be taken: Identify and quarantine any affected product found in inventory. If the product has been further distributed, please notify then of this recall. Firm is seeking product return. Questions about receiving credit for return merchandise, contact Jennifer Marsh at jmarsh@deroyal.com or by phone 865-362-4203 or email us at recalls@deroyal.com. If you have questions regarding this recall please contact us at 865-362-6465 or email us at recalls@deroyal.com.

More recalls by Deroyal Industries Inc

View all recalls by this company →