DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B
- Company
- Deroyal Industries Inc
- Recall Initiated
- February 9, 2023
- Posted
- March 23, 2023
- Recall Number
- Z-1283-2023
- Quantity
- 27,200 devices
- Firm Location
- Powell, TN
Reason for Recall
The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
Distribution
US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.
Lot / Code Info
GTIN 50749756377046, Lot 58447585 exp. 12/22/2027
Root Cause
Process control
Action Taken
The firm issued an URGENT DEROYAL RECALL NOTICE its consignees by email and letter beginning on 02/09/2023. The notice explained the problem and risk, and requested the following actions be taken: Identify and quarantine any affected product found in inventory. If the product has been further distributed, please notify then of this recall. Firm is seeking product return. Questions about receiving credit for return merchandise, contact Jennifer Marsh at jmarsh@deroyal.com or by phone 865-362-4203 or email us at recalls@deroyal.com. If you have questions regarding this recall please contact us at 865-362-6465 or email us at recalls@deroyal.com.