Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.

Company
Future Diagnostics Solutions B.V. Nieuweweg 279 Wijchen Netherlands
Recall Initiated
December 6, 2022
Posted
January 25, 2023
Recall Number
Z-1016-2023
Quantity
342 kits

Reason for Recall

Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.

Distribution

Worldwide distribution. US states of AZ, CA, CO, LA, OH, PA, and TX; Austria, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Spain, Sweden, and United Kingdom

Lot / Code Info

Catalog Number: 4K-IPT-00 UDI-DI: 08719326045214-230915 Lot Numbers: M78559 M78560 M78636

Root Cause

Under Investigation by firm

Action Taken

On 12/06/2022, the firm send a Recall Notice via email to customer informing them that The PTH kit 4K-IPT-00 with lot numbers: M78559, M78560 and M78636 can occasionally give a high duplicate %CV outside the recommended maximum of 15%CV. As a result of the failure in the duplicate CV requires a repeat run it may cause a delay in the patient result turnaround time or in system calibration depending on what samples are run and can lead to a wrong sample result. Actions to be taken by customers: -Immediately identify and quarantine the affected lot numbers. -Further instructions will be provided by the Future Diagnostics. For questions, send an email to fieldsafety@future-diagnostics.nl.