RecallDepth

Philips, REF M5071A, Adult. Smart PADS CARTRIDGE, 8+ years FOR USE WITH HEARSTART HS1 DEFIBRILLATORS

Company
Philips North America Llc
Recall Initiated
February 4, 2022
Posted
April 6, 2022
Recall Number
Z-0881-2022
Quantity
1,696,047
Firm Location
Bothell, WA

Reason for Recall

There is a potential for AED pads to experience gel separation from the foam/tin backing when peeled from the plastic liner. The gel may fold onto itself resulting in reduced surface area or gel on the pad, or it may separate almost completely leaving only a small amount of gel on the pad.

Distribution

Worldwide distribution. US Nationwide including Puerto Rico and Guam, Germany, Australia, Sweden, Japan, Canada, Taiwan, France, Netherlands, Spain, Singapore, Austria, South Africa, Italy, United Kingdom, India, Malaysia, Switzerland, Chile, Ireland, Norway, Hong Kong, Argentina, Mexico, Bolivia, Croatia, French Guiana, Denmark, United Arab Emirates, Hungary, Belgium, Indonesia, Korea, Republic of, Bhutan, New Zealand, Bulgaria, Mauritius, Trinidad and Tobago, Israel, Guatemala, Mali, Algeria, Philippines, Malta, Curacao, Panama, Paraguay, Costa Rica, Colombia, Liechtenstein, Saudi Arabia, Monaco, Estonia, Senegal, Reunion, Luxembourg, Qatar, El Salvador, Namibia, Myanmar, Nigeria, Poland, Aruba, Greece, Martinique, Guadeloupe, Dominican Republic, China, Kosovo, Portugal, Bangladesh, Tanzania, Peru, Lithuania, New Caledonia, State of Palestine, Finland, Cambodia, Macao, Cyprus, Zambia, Viet Nam, Uruguay, Ecuador, Netherlands Antilles, Czech Republic, French Polynesia, Kenya, Cameroon, Guinea, Mayotte, and Brunei Darussalam.

Lot / Code Info

ALL LOTS

Root Cause

Under Investigation by firm

Action Taken

On March 4, 2022, Philips mailed an Urgent Medical Device Correction notification. On December 14, 2022, Philips issued an Urgent Medical Device Correction - Expansion notification, dated November 15, 2022. The recall was expanded to notify customers who received an HS1/OnSite/Home AED more than 10 years ago. 1. Keep your HS1/OnSite/Home AED in service until you receive updated pads. If you need to use your HS1/OnSite/Home AED before receiving updated pads cartridge, ensure the majority of the pad surface is covered with gel and apply the pads to the patient. If the gel begins to separate from the white foam backing as you peel, try to prevent the gel from folding onto itself. Do not hesitate to apply the pads to the patient unless the gel has almost completely separated from the backing. 2. In case of trouble, install a spare pads cartridge and continue the rescue. No matter the state of the pads, follow the voice prompts as the HS1/OnSite/Home AED will guide you through the needed actions. If there is no spare pads cartridge, attend to the patient, providing CPR if needed, until EMS Personnel arrive. Store a spare pads cartridge with your HS1/OnSite/Home AED. 3. Unexpired M5071A Adult pads cartridges with a LOT number that begins with Y are eligible to be replaced with an updated Adult pads cartridge, free-of-charge. You must respond to receive any free-of-charge updated Adult pads cartridges. 4. Pass this notice to all those who need to be aware in your organization or to anywhere HS1/OnSite/Home AED devices or pads cartridges have been transferred. Keep a copy of this letter with the IFU/Owners Manual. 5. A similar notice was previously sent to customers who purchased HS1/OnSite/Home AEDs less than 10 years ago. If you received that notification, please respond to both notifications. For further information or support, contact your Philips representative.

More recalls by Philips North America Llc

View all recalls by this company →