Collagen Plugs, CP3, 0.3mm, CP4, 0.4mm, CP5, 0.5 mm, Sterile, EO, Rx Only, CE Dry eye treatment
- Company
- Lacrimedics Inc
- Recall Initiated
- January 17, 2022
- Recall Number
- Z-0709-2022
- Quantity
- 4780 units
- Firm Location
- Dupont, WA
Reason for Recall
The sterile pouch seal may contain channels that could affect the seal integrity and potentially compromise sterility.
Distribution
Nationwide and Worldwide Distribution to following countries - AR. BE, BO, BR, CH, CN, CO, CZ, DE, EC, ES, FI, GE, IN, IL, MY, MX, NZ, PE, RO, SG, SA, TR, TW, TZ, UK
Lot / Code Info
CP3 010421-3358 CP3 031419-3125 CP3 032921-3429 CP3 042020-3284 CP3 052219-3149 CP3 061521-3482 CP3 071720-3322 CP3 080218-3045 CP3 091218-3057 CP3 122118-3093 CP4 010621-3377 CP4 022619-3094 CP4 041620-3285 CP4 041719-3126 CP4 072020-3323 CP4 092518-3058 CP5 011221-3359 CP5 033121-3430 CP5 042219-3127 CP5 042220-3286 CP5 072320-3324 CP5 082318-3010 CP5 111418-3059 HP7 103018-3062
Root Cause
Packaging process control
Action Taken
Urgent : Medical Device Recall, Collagen Plugs and Opaque Herrick Lacrimal Plugs customer letters, dated January 17, 2022 and Urgent : Medical Device Recall, Collagen Plugs and Opaque Herrick Lacrimal Plugs distributor letters, dated January 17, 2022 sent starting 1/24/2022 (through 1/28/2022 US and 2/4/2022 (OUS)) via USPS certified letter with return receipt; Distributors and customers are instructed in letter to discontinue distribution and/or use and return remaining product for replacement " Contact their customers regarding the product(s) and request the return of remaining product for replacement. (distributors) " Acknowledge recall by returning the Recall Response Form to Tanya Traband: TTraband@lacrimedics.com or FAX: (253) 964-2699 or Lacrimedics, Inc. 2620 Williamson Place, Suite 113 DuPont, WA 98327 Other information: Contact Tanya Traband, Manager Client Relationships, at TTraband@lacrimedics.com or +1.800.367.8327 Ex. 200 Monday through Friday 8:30 a.m. to 4:00 p.m. PST. " Direct customers are instructed to return product for replacement; prepaid UPS labels are sent with the letter for this purpose