RecallDepth

Concerto User Interface Software provided with the following systems: (1) Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy System; (2) Xstrahl 150 Electronic Brachytherapy - 150kV Superficial X-Ray Therapy System; (3) Xstrahl 200 - 225kV Superficial / Orthovoltage X-Ray Therapy System; (4) Xstrahl 300 - 300kV Orthovoltage X-Ray Therapy System; (5) Xstrahl X 80 Photoelectric Therapy System.

Company
Xstrahl Limited
Recall Initiated
April 26, 2021
Terminated
January 26, 2024
Recall Number
Z-0174-2022
Quantity
25 systems in the U.S.

Reason for Recall

If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX. The countries of Austria, Germany, and Switzerland.

Lot / Code Info

V2.0, V2.1, and V2.2 when 2 opposing beam treatment plans are used.

Root Cause

Software design

Action Taken

The recalling firm issued an Urgent Field Safety Notice dated 4/26/2021 via email on 4/26/2021. The notice explained the issue, impact of the issue, and provided actions to be taken for Concerto 2.x 2 Beam Plans Only, although all customers with the affected software were issued the safety notice. The consignee was informed that until an update to their affected software version is installed, they should follow the actions listed in the letter.

More recalls by Xstrahl Limited

View all recalls by this company →