Atellica IM 1600 Analyzer, SMN 11066000

Company
Siemens Healthcare Diagnostics, Inc
Recall Initiated
September 28, 2021
Recall Number
Z-0125-2022
Quantity
1733
Firm Location
East Walpole, MA

Reason for Recall

Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result, customers at these altitudes may observe an increased number of Reagent Volume Check Errors.

Distribution

Domestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX UT VA WA WI WY. Foreign distribution worldwide.

Lot / Code Info

All units.

Root Cause

Under Investigation by firm

Action Taken

On September 27, 2021, the firm initiated notifications to affected customers. Customer distribution included all customers located in a city at an altitude of 350 meters (1148 feet) or higher. Customers were informed that Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350 (m) or 1148 (ft) above sea level. The letters provide directions for mitigating the potential for leakage. If you have any questions, contact your local Siemens Healthineers technical support representative.