Revolution CT

This recall has been completed (originally issued July 8, 2021).

Company
Ge Healthcare, Llc
Recall Initiated
March 19, 2021
Posted
July 8, 2021
Recall Number
Z-2004-2021
Quantity
197 (US = 40; OUS = 157)
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.

Distribution

Worldwide distribution - US Nationwide

Lot / Code Info

5590000 5590000-6 5590000-2

Root Cause

Radiation Control for Health and Safety Act

Action Taken

GE Healthcare will perform an initial site assessment and replace the collimator, if needed. This will be followed by bi-weekly remote monitoring of the system s log file. Finally, GE Healthcare will install a currently under development software patch, added to the daily fast calibration to automatically and reliably detect this failure should it occur. All of these actions will be performed at no cost to the customer. For questions call 262-544-3217.