Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.

Company
The Biomed Guys
Recall Initiated
March 25, 2021
Recall Number
Z-1651-2021
Quantity
22 units
Firm Location
Clarksville, TN

Reason for Recall

Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.

Distribution

Us distribution to the states of IN and TX.

Lot / Code Info

Serial numbers of serviced devices, serviced 09/2020-02-2021: 12833065, 13187021, 13187811, 12838069, 12982509, 12837231, 12835186, 12837972, 12710593, 12776790, 13753893, 13605919, 12832544, 15763526, 14806203, 14806612, 14807070, 13738432, 15763547, 15865626, 14806472, 12796415,

Root Cause

Under Investigation by firm

Action Taken

The firm notified their consignees via email 3/25/2021 and multiple phone calls 3/25/2021, 3/26/2021 & 3/30/2021. Customers where also sent empty boxes with packing material to return their affected devices for inspection and remediation.