RecallDepth

stryker Blueprint Mixed Reality Glenoid Digitizer, Catalog Number MRUE201, packaged in stryker Blueprint Mixed Reality Instrument Kit, Catalog Number MRUEGLN; Total Shoulder Arthroplasty

Company
Tornier S.A.S. 161 Rue Lavoisier Montbonnot-Saint-Martin France
Recall Initiated
March 5, 2025
Posted
April 8, 2025
Recall Number
Z-1535-2025
Quantity
22 units

Reason for Recall

Stryker identified an elevated complaint rate related to difficulties using the Blueprint Mixed Reality Guidance Instrumentation system is multifactorial, but in each instance prohibits the system from meeting its intended use. The most commonly reported complaints were related to impaired software functioning, impaired instrumentation functioning, and general difficulties using the product.

Distribution

US: TX, MN, AK, PA, MA, TN, ID, and France, Canada, and Ireland

Lot / Code Info

All lots manufactured through Oct 2024, Distributed 23-Aug-2023 through 31-Dec-2024, GTIN 03700434022933

Root Cause

Device Design

Action Taken

Stryker issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 03/12/2025 via email. The notice explained the issue and requested the following: "1. Immediately check your internal inventory to locate the product listed on the attached business reply form, remove them from their point of use, and isolate/quarantine the unit(s) to prevent accidental use. 2. Return the enclosed business reply form by email to confirm receipt of this notification/document product segregation. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. b. If you are a distributor, note that you are responsible for notifying your affected customers." For questions or concerns, please contact memphis.fieldaction@stryker.com. On 03/27/2025, Stryker issued an updated notice to include and additional product via email.