Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number: REF# 9524 (Small) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.

This recall is currently active, issued March 26, 2025. It was issued by Checkpoint Surgical Inc.

Company
Checkpoint Surgical Inc
Recall Initiated
February 24, 2025
Posted
March 26, 2025
Recall Number
Z-1466-2025
Quantity
135
Firm Location
Cleveland, OH
Official Source
View on FDA website ↗

Reason for Recall

Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.

Distribution

US Nationwide distribution.

Lot / Code Info

UDI-DI B13095240 Small Lot 4318 Expires 01Apr2026 Lot 4325 Expires 01Apr2026 Lot 4368 Expires 01Dec2026

Root Cause

Process control

Action Taken

The recalling firm alerted affected hospitals by visit on February 24, 2025. Customers were asked to check inventory for affected lot numbers and to return the affected inventory for replacement. Notifications were issued on February 26, 2025.