Abbot Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 29 mm, NVRO-29, Sterile
This recall is currently active, issued November 20, 2024. It was issued by Abbott Medical.
- Company
- Abbott Medical
- Recall Initiated
- October 17, 2024
- Posted
- November 20, 2024
- Recall Number
- Z-0493-2025
- Quantity
- 3 OUS
- Firm Location
- Saint Paul, MN
- Official Source
- View on FDA website ↗
Reason for Recall
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
Distribution
New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
Lot / Code Info
UDI-DI (GTIN): 5415067045775, Serial Numbers: 20150441, 20156804, & 20170535
Root Cause
Under Investigation by firm
Action Taken
A Medical Device Recall letter was hand delivered Oct. 21, 2024 to the one US consignee for 1 Impacted Valve. Nine foreign consignees were also notified for 26 Impacted Valves distributed outside of the United States (OUS). Customers are advised to return any remaining unused Impacted Valves, if applicable, and complete and return the accompanying Acknowledgement Form to Abbott. Abbott is executing further testing to verify that long-term durability of the Impacted Valves has not been affected and will communicate any changes to the recommendations set forth in this letter if needed by Q4 2025. In the interim, Abbott recommends continuing the same local standard of care in patients implanted with the Impacted Valve as would be used with patients implanted with any other transcatheter aortic valve implant. If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664 (Option 2).