Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00

Company
Spectrum Medical Ltd. Harrier 4, Meteor Business Park Cheltenham Road East Gloucester United Kingdom
Recall Initiated
March 13, 2023
Posted
October 24, 2024
Recall Number
Z-0224-2025
Quantity
172 units

Reason for Recall

The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.

Distribution

US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia.

Lot / Code Info

UDI/DI 05060434420770: Serial Numbers: a) AU6004040, AU6004041, AU6004042, AU6004043, AU6004044, AU6004046, AU6004050, AU6004051, AU6004052, AU6004053, AU6004054, AU6004056, AU6004057, AU6004058, AU6004060, AU6004061, AU6004062, AU6004063, AU6004064, AU6004065, AU6004066, AU6004069, AU6004070, AU6004071, AU6004072, AU6004073, AU6004074, AU6004075, AU6004136, AU6004137, AU6004138, AU6004139, AU6004140, AU6004141, AU6004142, AU6004143, AU6004144, AU6004145, AU6004146, AU6004147, AU6004148, AU6004159, AU6004160, AU6004161, AU6004162, AU6004167, AU6004168, AU6004169, AU6004170, AU6004171, AU6004172, AU6004173, AU6004174, AU6004175, AU6004176, AU6004177, AU6004178, AU6004179, AU6004180, AU6004181, AU6004182, AU6004183, AU6004184, AU6004185, AU6004186, AU6004187, AU6004188, AU6004195, AU6004197, AU6004198, AU6004200, AU6004201, AU6004202, AU6004203, AU6004204, AU6004205, AU6004207, AU6004332, AU6004333, AU6004348, AU6004349, AU6004350, AU6004352, AU6004353, AU6004354, AU6004532, AU6004577, AU6004578, AU6004579, AU6004580, AU6004587, AU6004588, AU6004590, AU6004591, AU6004603, AU6004604, AU6004605, AU6004606, AU6004607, AU6004609, AU6004610, AU6004613, AU6004731, AU6004771, AU6004773, AU6004776, AU6004777, AU6004778, AU6004779, AU6004781, AU6004782, AU6004919, AU6004920, AU6004921, AU6004922, AU6004923, AU6004924, AU6004925, AU6004926, AU6004927, AU6004929, AU6004930, AU6004955, AU6004957, AU6004958, AU6004960, AU6004963, AU6004966, AU6004968, AU6004970, AU6004971, AU6004972, AU6005010, AU6005021, AU6005022, AU6005023, AU6005024, AU6005025, AU6005026, AU6005027, AU6005028, AU6005029, AU6005030, AU6005037, AU6005038, AU6005039, AU6005040, AU6005041, AU6005042, AU6005057, AU6005065, AU6005109, AU6005110, AU6005118, AU6005124, AU6005125, AU6005355, AU6006351; Serial Numbers: b) AU8001273, AU8001275, AU8001346, AU8001347, AU8001348, AU8001359, AU8001360, AU8001366, AU8001403, AU8001406, AU8001407, AU8001446, AU8001447, AU8001449

Root Cause

Component design/selection

Action Taken

All potentially affected product was identified. The firm corrected all products between 03/13/2024 and 08/07/2024 during service visits.