Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.

This recall is currently active, issued October 2, 2024. It was issued by Mercury Enterprises, Inc. Dba Mercury Medical.

Company
Mercury Enterprises, Inc. Dba Mercury Medical
Recall Initiated
August 30, 2024
Posted
October 2, 2024
Recall Number
Z-0004-2025
Quantity
1,300 units
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

Manual resuscitator circuit flow controller may come apart due to a missing manufacturing step.

Distribution

US distribution to MI, FL, SC, PA, IN. International distribution to Albania.

Lot / Code Info

Part Number: 1050805; UDI/DI: 10641043508053, 30641043508057; Lot Numbers: 2417650805, 2417750805. Part Number: 1050832; UDI/DI: 10641043508329, 30641043508323; Lot Numbers: 2423550832, 2417850832.

Root Cause

Process control

Action Taken

Mercury Medical notified customers on 09/04/2024 via letter. Consignees who are medical facilities were instructed to immediately review inventory and quarantine any affected units on hand, contact customer service to arrange return of units, and complete the return response form. Consignees who are distributors were instructed to notify customers if further distributed, obtain a response form from those customers, arrange for any return of affected units, and complete the return response form.