Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22

This recall has been terminated (originally issued April 13, 2018).

Company
Trimed Inc.
Recall Initiated
April 13, 2018
Terminated
January 22, 2021
Recall Number
Z-2263-2018
Quantity
323
Firm Location
Santa Clarita, CA
Official Source
View on FDA website ↗

Reason for Recall

During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Distribution

U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden

Lot / Code Info

Lot #: a) R900, 009J, 050511, 00AG, 0095, 183601, 10505-11, 193601; b) R903, 26404, 264-04, 03NK, 050512, 10505-12, 24604, 11811, 181102, 181103, 188803; c) 26405, 009N, 00AL, R904, 00A1, 00PN, 050513, 10505-13, 183501, 83501, 264-05, 004L,

Root Cause

Process design

Action Taken

On 04/13/18, customer recall communication forms were sent via mail informing customers to quarantine effected devices, call the manufacturer to arrange device return at 800-633-7221, and report adverse events to the FDA's MedWatch.