Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment

This recall has been terminated (originally issued February 22, 2017).

Company
Terumo Cardiovascular Systems Corporation
Recall Initiated
January 5, 2015
Posted
February 22, 2017
Terminated
February 23, 2017
Recall Number
Z-1232-2017
Quantity
1463
Firm Location
Elkton, MD
Official Source
View on FDA website ↗

Reason for Recall

Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.

Distribution

The product was distributed domestically and also to foreign consignees. But not to government agency's or Canada.

Lot / Code Info

Catalog Number: 401-231, Product Description: Titan" Stabilizer Attachment , Lot Number: 052R   Catalog Number 401-231u , Product Description TitanTM 360 Stabilizer Attachment , Lot Number: 020R, 024R, 026R, 029R, 030R,034R,035R,037R, 040 , 048R

Root Cause

Other

Action Taken

A recall notification is being forwarded to all users and consignees who have received product. The notification of this voluntary recall will be sent to the following consignee in each facility:+ Director of Operating Room Services Proposed Level of Effectiveness Checks: Terumo CVS will verify that 100% of all letters are delivered to affected consignees