ProSun Onyx 32 Sli Intensive (12 minutes) tanning bed, Model 32/1. Labeled as the following: PROSUN ONYX 32 SLI INTENSIVE RED, PROSUN ONYX 32 SLI INTENSIVE PEARL WHITE, ANYTIME FITNESS ONYX 32 SLI INTENSIVE BRONZE, PROSUN ONYX 32 SLI INTENSIVE BRONZE Tanning of the skin

Company
Prosun International, Llc
Recall Initiated
December 10, 2018
Terminated
May 2, 2023
Recall Number
Z-1176-2019
Quantity
177 devices of which 31 are affected
Firm Location
Saint Petersburg, FL

Reason for Recall

ProSun International LLC discovered discrepancies in their Quality Control Checklist documentation, which could result in high UV output.

Distribution

US Nationwide distribution to states of: CA, CO, FL, GA, IL, KY, LA, ME, MI, NC, NJ, NY, OH, OR, TN, TX, and WI.

Lot / Code Info

Serial Numbers: UNO32SLII4W0048, UNO32SLII1X0052, UNO32SLII1X0063, UNATF32SLII1X0028, UNO32SLII1X0064, UNO32SLII1X0065, UNO32SLII1X0067, UNO32SLII1X0068, UNO32SLII2X0071, UNO32SLII2X0072, UNO32SLII3V0070, UNO32SLII2X0075, UNO32SLII1X0069, UNO32SLII2X0073, UNATF32SLII2X0031, UNATF32SLII2X0030, UNO32SLII2X0084, UNO32SLII2X0082, UNO32SLII2X0083, UNO32SLII2X0081, UNO32SLII2X0091, UNO32SLII4W0050, UNO32SLII4W0047, UNO32SLII3W0011, UNO32SLII2X0079, UNO32SLII3X0095, UNO32SLII2X0094, UNO32SLII1X0056, UNO32SLII1X0057, UNATF32SLIIX0027, UNO32SLII3X0096

Root Cause

Employee error

Action Taken

ProSun initially notified customers on about 12/10/2018 via "URGENT: POTENTIAL UV EXPOSURE DISCREPANCY" letter. Instructions included to contact ProSun Customer Service at 727-825-0400 Ext 243 or 224 or by email at jenniferh@prosun.com to coordinate receipt of a UV meter to verify UV output. Once received, customers should take the requested UV readings, fill out the form accompanying the UV meter, and return the UV meter with signed form via the enclosed return label. If there are any out-of-specification readings present, ProSun will contact that customer with a solution. Another customer notification was sent to customers on about 3/11/19 via "URGENT: Urgent Recall Notice" letter. Instructions included were the same as the 12/10/18 letter. Customers who were initially contacted via telephone call received a follow up letter with the same instructions. If you have any questions, Contact ProSun Customer Service at 727-825-0400 Ext 224.