Presource Kit Total Joint Pack containing a double-wrapped basin component. Product Usage: This kit is used on patients undergoing joint replacement knee or hip.

Company
Cardinal Health Inc.
Recall Initiated
November 16, 2018
Terminated
November 4, 2019
Recall Number
Z-0702-2019
Quantity
182 packs
Firm Location
Dublin, OH

Reason for Recall

The double wrapping of this custom kit basin set is not supported by the current sterility assurance validation.

Distribution

US state of Florida

Lot / Code Info

Product code: SOP22TJCLK Lot numbers # 104405, 122793, and 987948

Root Cause

Unknown/Undetermined by firm

Action Taken

As only one customer is potentially impacted, in the interest of prompt action and notification, a teleconference was held on Friday, November 16, 2018 with the customer to communicate the impending actions. During this teleconference, the customer was informed that an HHE and field corrective action was in-process and that a study was being initiated to determine that their requested packaging configuration was acceptable. The customer also was informed that they should not continue to use the impacted double wrapped basin, and the components contained within, until the study was completed. Cardinal Health is following up on this teleconference with a written notification.