Optilite IgG4 Kit Product Code: LK009.OPT.A

This recall has been terminated (originally issued December 23, 2016).

Company
The Binding Site Group, Ltd.
Recall Initiated
December 23, 2016
Terminated
May 3, 2017
Recall Number
Z-1089-2017
Quantity
9 users
Firm Location
Birmingham United Kingdom
Official Source
View on FDA website ↗

Reason for Recall

A change to the lower end of the measuring range and product insert did not reflect the change.The measuring range provided in instrument parameters is correct but contradicts the information provided in the product insert

Distribution

U.S. distribution to the following; FL, TX, PA, CA, MA, WA, MI, CT. No foreign distribution.

Lot / Code Info

Lot #401413

Root Cause

Under Investigation by firm

Action Taken

Customer notification letters were sent on 12/23/16. The Binding Site recommends to continue testing on the kit as normal and refer to the measuring ranges and limit of quantification stated in the notice and as indicated in the amended insert (INS009.OPT.A, Version: 21st December 2016) provided with this notification. The notice needs to be passed on to all those who need to be aware within the organization or to any Organization where the potentially affected devices have been transferred. Return the completed E-Back Form to your local Binding Site Representative for records and note that The Binding Site intends to report this correction to the FDA recall coordinator. If you have any questions contact Technical Services.