RecallDepth

Human Lambda Free SPAPlus Kit Product Code: LK018.S Freelite aids in the detection and monitoring of Multiple Myeloma and related diseases.

This recall has been terminated (originally issued January 19, 2017).

Company
The Binding Site Group, Ltd.
Recall Initiated
November 28, 2012
Posted
January 19, 2017
Terminated
March 3, 2017
Recall Number
Z-1050-2017
Quantity
1,002
Firm Location
Birmingham United Kingdom
Official Source
View on FDA website ↗

Reason for Recall

Calibration curve activity has increased over time in the kit lots listed.

Distribution

U.S. distribution to the following; CA, TX, HI, GA, PA, UT, NY, OK, ID, MA, AZ. Foreign distribution to the following; Australia, Bulgaria, Denmark, Hong Kong, Ireland, Israel, Japan, New Zealand, Taiwan, Belgium, France, Italy, Spain.

Lot / Code Info

Lot #323278, 327963

Root Cause

Unknown/Undetermined by firm

Action Taken

Customer notification letters were sent 11/28/12. Customers were asked to discontinue using the listed kit lots. Discard any unused or partially-used kits of these lots. Let The Binding Site know how many replacement kits are required.

More recalls by The Binding Site Group, Ltd.

View all recalls by this company →