LeverEdge Contrast Injector: 35 cc syringe with ISO 594 female luer lock - Product Usage: intended to be used to perfuse contrast medias into the vessels for angiographic procedures, Catalog #: 4100-00.

Company
Lemaitre Vascular, Inc.
Recall Initiated
April 6, 2021
Terminated
December 29, 2022
Recall Number
Z-1529-2021
Quantity
616 units
Firm Location
Burlington, MA

Reason for Recall

The firm received a complaint regarding particulate found in the packaging foam that could contaminate the sterile field.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, CA, FL, IL, IN, KY, MA, MO, NY, OH, TN, TX and the countries of Bahamas, Belgium, France, Germany, Hong Kong, Spain.

Lot / Code Info

Lot: LCI1091, LCI1092, LCI1093, LCI1094, LCI1095, LCI1096, LCI1097, LCI1098, LCI1099

Root Cause

Process control

Action Taken

On 04/01/2021, LeMaitre issued an Urgent Medical Device Removal notice via letter informing customers that particulate contamination has been detected from the packaging foam with these devices.