ELI 280 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.
- Company
- Welch Allyn Inc Mortara
- Recall Initiated
- March 31, 2021
- Terminated
- August 16, 2024
- Recall Number
- Z-1482-2021
- Quantity
- 99 units
- Firm Location
- Milwaukee, WI
Reason for Recall
Devices do not meet IEC 60601-2-27 requirements as labeled.
Distribution
US Nationwide distribution in the states of CA, CO, FL, GA, IN, LA, NC, NJ, NV, NY, OH, OK, TX, UT, and VA.
Lot / Code Info
Part Number: ELI280-LDB-ADAAX (Device Identifier: 732094265699); S/N: 117030550918 and 117030550919 Part Number: ELI280-LDD-AAABX (Device Identifier: 732094265682); S/N: 117260576403 Part Number: ELI280-LDX-ADABX (Device Identifier: 732094265675); S/N: 118240001697 and 118150627678 Part Number: ELI280-LDX-ADFBD (Device Identifier: 732094265644); 118230000580, 118230000570, 118230000584, 118230000569, 118230000572 118230000582, 118230000574, 118230000576, 118230000577, 118230000581, 118230000568, 118230000575, 118230000583, 118230000578, 118230000571, 118230000567, 118340001621, 118340001622, 118340001623, 118400001475, 118400001471, 118400001472, 118400001476, 118400001467, 118400001470, 118400001469, 118400001482, 118400001477, 118400001481, 118400001478, 118400001479, 118400001480, 118420000070, 118420000071, 118420000065, 118420000067, 118420000068, 118420000066, 118420000069, 118430000923, 118430000946, 118430000930, 118430000932, 118430000928, 119150000011, 119150000001, 119150000002, 119150000003, 119150000010, 119150000008, 119150000009, 119150000004, 119150000007, 119180000631, 119180000632, 119180000633, 119180001155, 119180001156, 119180001153, 119180001154, 119180001157, 119260000043, 119260000442, 119260000443, 119280001568, 119270000064, 119290001054, 119290001053, 119290001048, 119290001052, 119290001050, 119290001049, 119290001051, 119290001055, 119470001878, 119470001879, 119510000132, 119510000135, 119510000134, 119510000133, 120080000860, 120050001405, 118120622367, 118180633837, 118180633838, 118180633839, 118180633840, 118190635342, 118190635353 Part Number: ELI280-LDX-ADFBG (Device Identifier: 732094319941); S/N: 119480000195, 119480000191, 119480000193, 119480000194, 119480000192
Root Cause
Process control
Action Taken
Recall notification letters were mailed to customers notifying them of the potential issue with recommended actions beginning April 7, 2021. The notification letter includes an acknowledgement form that must be returned to the recalling firm. Once the customer confirms the impacted units are in possession, the recalling firm will arrange for an exchange AM12M to be provided. The recalled unit will be returned to the recalling firm's service center for processing.