Tearaway Introducers, Catalog #: a) BNS-000-07, b) BNS-000-10; c) DIL-015-50; d) DSS-005-11; e) DSS-005-12; f) DSS-006-045; g) DSS-007-05; h) DSS-007-165; i) DSS-010-03; j) DSS-010-04; k) DSS-010-045; l) DSS-010-05; m) DSS-012-07; n) SPT-003-46

This recall has been terminated (originally issued May 2, 2018).

Company
Galt Medical Corporation
Recall Initiated
May 2, 2018
Terminated
July 20, 2020
Recall Number
Z-2565-2018
Quantity
23777
Firm Location
Garland, TX
Official Source
View on FDA website ↗

Reason for Recall

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

Distribution

U.S.: MN, PA, VA, CA, UT, FL, NY, GA, LA, MO, TX, MS, OH, WA, MA, IL, KY, NC, and Puerto Rico; OUS: Canada, Australia, France, Germany, Ireland, Italy, Netherlands, Poland, Slovakia, Switzerland, Belgium, United Kingdom, Korea

Lot / Code Info

Lot #: a) 18024880, 18043086, 18066833, 18073282; b) 18039555, 18059249, 18074835;  c) 17363134;  d) 17352818, 18051601; e) 18019228, 18047509; f) 18023833; g)   17356436, 18018694, 18018695, 18040074; h) 18061579; i) 18015027; j) 18019038, 18029067, 18033068, 18075816, 18075817, 18075818; k) 17363816, 18023834, 18026040, 18026072, 18033073, 18033079, 18054248; l) 18005018, 18012032, 18012034, 18012035, 18012036, 18019037, 18026953, 18089355, 18096357, 18096358, 18103359; m) 18047202;  n) 18054349

Root Cause

Other

Action Taken

On approximately 05/01/2018 recall letters and customer response forms were mailed to all consignees.