MA3320-PM Premium Adult - SYK Mistral-Air Premium Blankets. MODEL #: MA3302-PM. CATALOG #(s): MA3302-PM
- Recall Initiated
- July 21, 2017
- Terminated
- March 22, 2018
- Recall Number
- Z-0419-2018
- Quantity
- 11160
- Firm Location
- Portage, MI
Reason for Recall
It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).
Distribution
U.S.
Lot / Code Info
LOT #(s): 1640002, 1645003
Root Cause
Component design/selection
Action Taken
Stryker notified customers of the recall by sending Customer Notification and Business Reply Forms, dated July 21, 2017.