Captivator Single-Use Polypectomy Snares, Captiflex Single-Use Polypectomy Snares. Outer box UPN, Inner Pouch UPN: M00561311 (Box 10), M00561310 M00562301 (Box 10), M00562300; M00562321 (Box 10), M00562320; M00562341 (Box 10), M00562340; M00562451 (Box 10), M00562450; M00562401 (Box 10), M00562400; M00562402 (Box 40), M00562400; M00562422 (Box 40), M00562420; M00562471 (Box 10), M00562470.

This recall is currently active, issued January 19, 2021. It was issued by Boston Scientific Corporation.

Company
Boston Scientific Corporation
Recall Initiated
January 19, 2021
Recall Number
Z-1168-2021
Quantity
335,788 units total (Domestic: 222,882, Foreign: 112,906)
Firm Location
Marlborough, MA
Official Source
View on FDA website ↗

Reason for Recall

Packaging Incomplete Seal. Users noted an incomplete pouch seal prior to using the device in a patient.

Distribution

US nationwide distribution.

Lot / Code Info

Outer box GTIN: 08714729019251 Inner pouch GTIN: 08714729747680 Lot/Batch#: 24725319, 24782695,   Outer box GTIN: 08714729019312 Inner pouch GTIN: 08714729747833 Lot/Batch#: 24842597, 24812577,   Outer box GTIN: 08714729019336 Inner pouch GTIN: 08714729747857 Lot/Batch#: 24889757, 24719202, 24709536, 24773287,   Outer box GTIN: 08714729019350 Inner pouch GTIN: 08714729747871 Lot/Batch#: 24782693, 24732964, 24889755, 24780939,   Outer box GTIN: 08714729071068 Inner pouch GTIN: 08714729747970 Lot/Batch#: 24882357, 24782689, 24782687, 24729758,   Outer box GTIN: 08714729019411 Inner pouch GTIN:08714729747932 Lot/Batch#: 24961643  Outer box GTIN:08714729501640 Inner pouch GTIN: 08714729747932 Lot/Batch#: 24961645, 24989260, 24844193,   Outer box GTIN:08714729501657 Inner pouch GTIN:08714729747956 Lot/Batch#: 25010054,   Outer box GTIN:08714729019459 Inner pouch GTIN:08714729747994 Lot/Batch#: 25016793.

Root Cause

Process change control

Action Taken

On 01/21/2021, Boston Scientific issued Urgent Medical Device Product Removal notices to customers via e-mail. The removal notice is due to an incomplete seal of the inner pouch which could result in a sterile barrier breach rendering the specified lots of polypectomy snare devices non-sterile.