RecallDepth

OSTEORAPTOR 2.3 W 1UB II BLUE-Absorbable Suture anchor Product Number: 72201991

Company
Smith & Nephew, Inc.
Recall Initiated
September 9, 2020
Recall Number
Z-0148-2021
Quantity
263,874 (total)
Firm Location
Andover, MA

Reason for Recall

Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch

Distribution

Nationwide Foreign: AE, AT, AU, BE, BM, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GE, GR, HK, HU, IE, IL, IN, IT, JP,KR, LV,MY, NL, NO, PA,PE, PL, PR, PT, RU, SE, SG, SI, TH, TR, TW, UA, & ZA

Lot / Code Info

Batch Numbers: 2012767 2013997 2016120 2017772 2018792 2019839 2021494 2022257 2024709 2025285 2027451 2027481 2030907 2030908 2031041 2033235 2033236 2034376 2034377 2034378 2035450 2036270 2049647 2049648 2053756 2049649 2050880 2050881 2056052 2056053

Root Cause

Equipment maintenance

Action Taken

Smith & Nephew issued Urgent Medical Device Recall letter (R-2020-19) FSN on September 9, 2020 via email or overnight mail to customers that received affected products. Letter states reason for recall, health risk, and action to take: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.