DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US
- Recall Initiated
- June 30, 2020
- Terminated
- November 17, 2020
- Recall Number
- Z-2747-2020
- Quantity
- 1800 units
- Firm Location
- Newark, DE
Reason for Recall
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.
Distribution
Nationwide
Lot / Code Info
DCA Vantage Handheld Barcode Scanner #11306432 Model: DS4308 - HC0015BZZWW
Root Cause
Labeling design
Action Taken
Siemens Healthcare Diagnostics issues a field safety notice date June 2020 to customers