RecallDepth

DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US

Company
Siemens Healthcare Diagnostics, Inc.
Recall Initiated
June 30, 2020
Terminated
November 17, 2020
Recall Number
Z-2747-2020
Quantity
1800 units
Firm Location
Newark, DE

Reason for Recall

If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.

Distribution

Nationwide

Lot / Code Info

DCA Vantage Handheld Barcode Scanner #11306432 Model: DS4308 - HC0015BZZWW

Root Cause

Labeling design

Action Taken

Siemens Healthcare Diagnostics issues a field safety notice date June 2020 to customers

More recalls by Siemens Healthcare Diagnostics, Inc.

View all recalls by this company →