Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com

Company
Quidel Cardiovascular Inc
Recall Initiated
April 9, 2019
Terminated
January 14, 2021
Recall Number
Z-1552-2020
Quantity
123 kits
Firm Location
San Diego, CA

Reason for Recall

Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

Distribution

US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL

Lot / Code Info

Lot Number: C3470AN Part Number 94413

Root Cause

Software Manufacturing/Software Deployment

Action Taken

On April 10, 2019 Quidel issued a recall notice to affected consignees to take the following actions: 1. Check the Quidel Triage TOX Drug Screen Control 1you have on hand to determine if you have any of the affected lot number (C3470AN). 2. If you locate the above-noted lot number, please properly destroy them to ensure they are not used. 3. If you have forwarded this lot to another facility, please forward this communication to that facility. 4. Complete the enclosed Inventory Assessment/Certificate of Destruction form and return to CustomerNotifications@quidel.com. 5. If you have any questions or require assistance, please contact Quidel Technical Support: " For US and Canada, please call 800-874-1517, option 2 " For Europe and Middle East, please call +353 {91) 412 474