ImplantDirect InterActive Precision IO Scan Adapter, Part #6534-09PT, Platform: 3.4mm, Non-Sterile, Rx. The firm name on the label is Implant Direct Sybron Manufacturing LLC, Thousand Oaks, CA.

This recall has been completed (originally issued November 28, 2018).

Company
Implant Direct Sybron Manufacturing, Llc
Recall Initiated
November 28, 2018
Recall Number
Z-0790-2020
Quantity
274 devices
Firm Location
Thousand Oaks, CA
Official Source
View on FDA website ↗

Reason for Recall

The peek part was loose on the titanium base of the device which could result in it falling into a patient's mouth during a procedure resulting in aspiration or ingestion.

Distribution

Distribution was made to AL, AZ, CA, FL, GA, KS, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, and VA. There was no government/military distribution. Foreign distribution was made to Canada, Armenia, Australia, Denmark, France, Germany, Great Britain, Iceland, Italy, Lithuania, Netherlands, Poland, Spain, Sweden, and Switzerland.

Lot / Code Info

Lot numbers 104479 and 104963.

Root Cause

Process control

Action Taken

The firm first issued recall letters on 11/28/2018 via FedEx for Part number 6534-09PT and then on 12/28/2018, issued recall letters for Part number 6530-09PT via FedEx. On 2/14/2019, the firm issued another letter to their consignees that added a risk statement regarding the recall of both part numbers (6534-09PT and 6530-09PT).