RecallDepth

Semi-Rigid Suction Canister, 3000cc. Model Numbers OR530 and OR530H. For OR530, the product comes in a case with two boxes. One box includes 50 eaches of liners in a poly bag and the second box includes 50 eaches of lids in a poly bag. For OR530H, the customer receives the number of liners and lids purchased in one case without the individual boxes. The "H" denotes a single quantity when being ordered.

Company
Medline Industries Inc
Recall Initiated
January 31, 2019
Recall Number
Z-0174-2020
Quantity
750 eaches
Firm Location
Northfield, IL

Reason for Recall

Potential for the Semi-Rigid canister lid to fragment during use.

Distribution

Nationwide distribution to AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MT, NC, NH, NJ, NV, NY, OH, PA, RI, SC, TN, TX, VA, VT, WI.

Lot / Code Info

Model Numbers OR530, OR530H; Lot Numbers 20180914-1, 20180913-1, 20180916-1, 20180929-1

Root Cause

Process control

Action Taken

On about 01/31/2018 Medline notified customers via "URGENT RECALL" letter. Instructions included to immediately check inventory for and quarantine all affected product, complete and return the response form, and notify customers if the product has been further distributed and instruct them to return products to Medline Industries, Inc. Once the response form is received, return labels will be provided to return all affected product. For questions call 866-359-1704.

More recalls by Medline Industries Inc

View all recalls by this company →