RecallDepth

Rosebud Vital Signs Monitor with NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5 - Product Usage: The monitor is intended for spot-check monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), and Temperature (TEMP) on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians

Company
Mindray Ds Usa, Inc. D.B.A. Mindray North America
Recall Initiated
October 1, 2019
Terminated
August 6, 2021
Recall Number
Z-0238-2020
Quantity
9 units
Firm Location
Mahwah, NJ

Reason for Recall

NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function

Distribution

US Nationwide distributions.

Lot / Code Info

NIBP valve assembly board, p/n 115-017679-00, containing software version 8.1.5

Root Cause

No Marketing Application

Action Taken

Mindray communicated by letter on 10/1/19 to consignees and distributors by certified postal mail with return receipt verification: to consignees that have affected vital sign monitor; communication to consignees that purchased directly NIBP valve assembly boards; communication to distributors that have affected vital sign monitor. Consignees were instructed to contact Mindrays Repair Center to arrange for affected units to be returned for a software update.