Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.

Company
Seiler Instr. & Mfg. Co. Inc.
Recall Initiated
January 1, 2017
Recall Number
Z-1294-2018
Quantity
43 units, but 1 affected
Firm Location
Saint Louis, MO

Reason for Recall

The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.

Distribution

Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.

Lot / Code Info

Model Number 6 ENT

Root Cause

Use error

Action Taken

The customer was notified via email on about January 4, 2017. Instructions included that the scope only tips if the user has it extended all the way out and is pulling on the handles of the scope. The end user must not pull on the handles of the scope when it is extended. They were also instructed to educate personnel on this matter. For further questions, please call (314) 968-2282.