Excella II Rod Cat. No.: SR 55100 5.5 x 100 mm LOT 1502 and Excella II Rod Cat. No.: SR 55110 5.5 x 110 mm LOT 1502 Packaged in heat-sealed plastic bag. Also provided in reusable steam sterilization tray as implant set. Intended for use in the non-cervical area of the spine.

This recall has been terminated (originally issued January 19, 2016).

Company
Innovasis, Inc
Recall Initiated
December 7, 2015
Posted
January 19, 2016
Terminated
April 7, 2016
Recall Number
Z-0662-2016
Quantity
17 units Labeled SR55100 Lot 1502 with SR55110 rods in package
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

Innovasis recalled two lots of Excella II titanium rods which were packaged using each other's labels.

Distribution

Distributed in US -- CA; KS; OH; & TX.

Lot / Code Info

Catalog No.: SR55110 Lot 1502 and SR55100 Lot 1502

Root Cause

Labeling mix-ups

Action Taken

Innovasis sent a Voluntary Product Recall letter, dated December 7, 2015 to customers regarding two lots of Excella II titanium rods which were packaged using each other's labels. Customers were asked to check their inventory and either remove the rod from the mislabeled package and request additional rods, or return the unopened packages immediately. The firm will replace all returned inventory upon receipt.