ARIA Radiation Oncology, versions 10, 11 , 13.0 and 13.5 with Clinical Assessment License. Model number HIT Product Usage: ARIA Radiation Oncology is a radiation treatment plan and image management application.

This recall has been terminated (originally issued October 30, 2015).

Company
Varian Medical Systems, Inc.
Recall Initiated
September 28, 2015
Posted
October 30, 2015
Terminated
May 12, 2017
Recall Number
Z-0170-2016
Quantity
1971 units
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

An anomaly was identified with the ARIA for Radiation Oncology software with a Clinical Assessment license. Admin Instructions may not print on drug order prescriptions or be transmitted with e-Rx.

Distribution

Worldwide Distribution - US Nationwide and the countries of: Canada, Switzerland, South Korea, Germany, Spain, Ireland, Spain, France, Belgium, United Arab Emirates, United Kingdom, Norway, Italy, Trinidad and Tobago, Netherlands, Philippines, Portugal, Denmark, Sweden, Belgium, Saudi Arabia, Thailand, Australia, India, Taiwan, South Africa, Israel, Australia, Kuwait, Finland, Tunisia, Algeria, Argentina, Bangladesh, Brazil, Chile, China, Colombia, Costa Rica, Czech Republic, Guam, Guatemala, Hong Kong, Indonesia, Japan, Kazakhstan, Kuwait, Lebanon, Luxembourg, Madagascar, Moldova, Morocco, New Zealand, Panama, Poland, Qatar, Romania, Russia, Singapore, South Africa, Tunisia and Turkey.

Lot / Code Info

Model number HIT: All units with software version 10, 11, 13.0 and 13.5 with Clinical Assessment license.

Root Cause

Software design (manufacturing process)

Action Taken

Varian sent an Urgent Medical Device Corrections letter dated September 29, 2015 via trackable method to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Varian is developing a technical correction for this issue. Customers will be contacted by a Varian Field Service. A copy of the notification is to be kept with the most current product labeling and all appropriate personnel in the radiation oncology department should be aware of this issue. For questions contact your local Varian Medical Systems Customer Support District or Regional Manager