Metasul Femoral Head - A modular head component is used in conjunction with a femoral stem component for replacement of the proximal femur in total hip arthroplasty.

This recall has been terminated (originally issued February 20, 2015).

Recall Initiated
December 17, 2014
Posted
February 20, 2015
Terminated
September 23, 2015
Recall Number
Z-1133-2015
Quantity
30
Official Source
View on FDA website ↗

Reason for Recall

Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.

Distribution

Domestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY International: Canada, Japan

Lot / Code Info

Item numbers: 19.32.05; 19.32.06; 19.32.08;  Lot #'s: 2560585, 2553420, 2556039, 2560586, 2563398

Root Cause

Package design/selection

Action Taken

On 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.