RecallDepth

Timberline Anchored Lateral Fixed Retractable Sleeve Model No. 8630-0205. Timberline Anchored Lateral Fixed Retractable Sleeves are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae. The Timberline Retractable Drill and the Timberline Retractable Awl fit within this Sleeve. Spinal Surgical Device.

This recall has been terminated (originally issued January 15, 2015).

Company
Biomet Spine Llc.
Recall Initiated
August 22, 2013
Posted
January 15, 2015
Terminated
March 13, 2015
Recall Number
Z-0983-2015
Quantity
14 units
Firm Location
Broomfield, CO
Official Source
View on FDA website ↗

Reason for Recall

The inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve.

Distribution

US Distribution to states of: AZ, CA, NY, TN and TX including PR.

Lot / Code Info

LOT Numbers: TU00053A, TU00053B and TU00118 Product Code: 86300205

Root Cause

Device Design

Action Taken

The firm, Lanx, sent a letter to their Consignees/customers dated October 15, 2013. The letter describes the product, problem and actions taken. Lanx and their distributors replaced the affected inventory with conforming product. Please contact Chief Compliance Officer at 303-501-8534 or by email: Randy.fesmire@lanx.com if you have any questions regarding this recall.