Terumo Custom Cardiovascular and Perfusion Procedure Kits Catalog Codes: Kits with a 5-digit code beginning with the digit 6: Catalog Number: 6XXXX ; Catalog Codes: Kits with a 5-digit code beginning with the digit 7: Catalog Number: 7XXXX Product usage: The Cardioplegia Delivery Sets are indicated for delivery of cardioplegic solutions to the heart during cardiopulmonary bypass surgery. The Cardiovascular Procedure Kits are indicated for use only in the extracorporeal circuit for the cardiopulmonary bypass procedure(s) for which the user designed it.

This recall has been terminated (originally issued October 31, 2014).

Company
Terumo Cardiovascular Systems Corp
Recall Initiated
March 1, 2011
Posted
October 31, 2014
Terminated
November 6, 2014
Recall Number
Z-0159-2015
Quantity
Catalog Number: 6XXXX: 4561 units; Catalog Number: 7xxxx : 19356 units
Firm Location
Ashland, MA
Official Source
View on FDA website ↗

Reason for Recall

Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the table line.

Distribution

Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand.

Lot / Code Info

Catalog Number: 6XXXX Lot Numbers: ML07, MM18, MM25, MN08, MN IS, MN22.  Manufacturing dates September 7, 20 I 0 through November 22, 20 I 0.  Catalog Number: 7XXXX Lot Numbers: ML07,ML20, MM18, MM25, MNO1, MN08, MN15, MN29,MP06, MP13. Manufacturing dates September 7, 2010 through December 13,2010.

Root Cause

Device Design

Action Taken

Terumo Cardiovascualr notified accounts by letter dated 3/1/2011 advising users users that the luer connector to the table line in certain lots of myocardial protection sets may leak. Users should not continue to use affected myocardial protection sets in a manner that exposes the connector to negative pressure without taking correction or replacement activities as described. Firm provided mitigating instructions: Clamp the table line distal to the luer connector when ending an infusion of cardioplegia thus isolating the leak from negative pressure. Contact Terumo CVS Customer Service to report the defect and arrange return of the unit: 800-521-2818.

More recalls by Terumo Cardiovascular Systems Corp