Hill-Rom 1000 Bed, Product No. P1160. The Hill-Rom 1000 Bed is intended for low to moderate acuity patient in the medical surgical area of the hospital.

This recall has been terminated (originally issued June 5, 2013).

Company
Hill-Rom, Inc.
Recall Initiated
April 15, 2013
Posted
June 5, 2013
Terminated
February 7, 2017
Recall Number
Z-1466-2013
Quantity
8,737 devices
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

Complaints have been reported for brakes disengaging when the bed is subjected to a hard jolt or force to the frame.

Distribution

Worldwide distribution: US (Nationwide) and countries of: Argentina , Bahamas, Bahrain, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, Ethiopia, France, Ghana, Guam, Hong Kong, India, Indonesia, Japan, Jordan, Kuwait, Lebanon, Libya, Malaysia, Mexico, N. Mariana Islands, Pakistan, panama, Paraguay, Puerto Rico, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, United Arab Emirates, United States, US Virgin Islands, Venezuela and Vietnam.

Lot / Code Info

Product number (P1160) with Serial Numbers J143AP0937 thru O080AP5661.

Root Cause

Process control

Action Taken

On 4/16/2013 an URGENT MEDICAL DEVICE CORRECTION notification letter was sent to all consignees. The firm plans a visit to all facilities to correct each bed. Call Hill-Rom Technical Support at 800-445-3720 if you have any questions.