RecallDepth

enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.

This recall has been terminated (originally issued July 9, 2012).

Company
Vital Signs Devices, A Ge Healthcare Company
Recall Initiated
February 20, 2012
Posted
July 9, 2012
Terminated
August 19, 2013
Recall Number
Z-1972-2012
Quantity
11803 cases
Firm Location
Totowa, NJ
Official Source
View on FDA website ↗

Reason for Recall

The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.

Distribution

Nationwide Distribution.

Lot / Code Info

a) 510(k): K060537 b) Medical device listing number: D071942  All enFlow Warmer Straps in the field present a potential for tissue injury. Sold separately as Product #: 980304 and 980304EU Included as an accessory in following enFlow products:  Military Domestic 91000153 980100 91000154 980100EU 91000156 980105VS 91000103

Root Cause

Device Design

Action Taken

GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included customer instructions and a Customer Response Form to be completed and returned to the firm. Contact Customer Service at 1-800-932-0760 for questions regarding this recall.

More recalls by Vital Signs Devices, A Ge Healthcare Company